MODEL 2121F CHARGE AMPLIFIER & MODEL 4121 POWER SUPPLY

2121f

BOUCHE LABS MODEL 2121F CHARGE AMPLIFIER & MODEL 4121 POWER SUPPLY

BOUCHE LABS Model 2121F Charge Amplifier is low-priced and is used with piezoelectric transducers. The small overall dimensions and DC-power supply or battery operation is appropriate for use in the test laboratory or in the field.


An easy two step process is used for operation:

  • First, dial the transducer sensitivity (1 to 1100 pC/g)
  • Second, select the gain switch position (1 to 10,000 mV/g.) Then the charge amplifier output is in mV/g.


Another attribute is the front panel switch for selecting a single-pole low-pass filter for suppressing resonances during shock tests.

The 2121F Charge Amplifier is supplied with +15 Vdc and -15 Vdc by the Bouche Labs Model 4121 AC-operated power supply. The power supply is capable of powering up to twenty Model 2120 series Bouche Labs charge amplifiers.

Specifications

Measurement Unit Rating
INPUT 12
MODE Single-ended
GROUND (Insulated from case) ohms 10M
TRANSFER +/-2
SYSTEM SENSITIVITY The gain is adjustable to provide output for transducer sensitivities from 1 to 1100 pC/g
Transducer sensitivity dial pC/g 100-1100, 10-110, 1-11
Amplifier gain mV/g 100 to 10000, 10 to 1000, 1 to 100
ACCURACY (on any range) % +/-0.3
GAIN STABILITY maximum per 1000 pF change in source capacitance % 0.1
FREQUENCY RESPONSE, ref: 100 Hz
Filter Out: 10 Hz to 10000 Hz % +/-0.5
...at 5 Hz: % +0.2/-1.0
Filter in: 5 Hz to 15 kHz % +/-5
...at 22 kHz, minimum dB -3
...at 28 kHz, minimum dB -10
Terminal slope dB/Oct -18
AMPLITUDE LINEARITY (ref:best straight line) % +/-0.2
WARM-UP TIME min 1
RESIDUAL NOISE (referred to input) pC 0.3 (rms)
TRANSIENT DISTORTION
maximum time delay optional low-pass filter, any frequency 5 Hz to 15 kHz us 4
TRANSIENT DISTORTION
maximum time delay of optional low-pass filter, any frequency 5 Hz to 15 kHz us 4
POWER 1/4-28, 10-32, etc.
LINE VOLTAGE Vac 105 to 125
CIRCUIT GROUND, Insulated from case, minimum ohm 10M
Characteristic Unit Rating
POWER REQUIREMENTS
SUPPLY VOLTAGE Vdc +15/-15, typical limits +14/-14 to +18/-18
SUPPLY CURRENT mA 7.5, typical, 9.5 maximum, at +15/-15 Vdc
PHYSICAL 3
OVERALL DIMENSIONS in/cm 1.75 x 4.5 x 6.25/4.4 x 11.4 x 15.9
FINISH Anodized Aluminum
WEIGHT lb 2
CONNECTORS
Input 10-32 Microdot S-50 or equivalent
Output BNC, UG-1094/U or equivalent
ENVIRONMENTAL 95
OPERATING TEMPERATURE oF/oC +40 to +125/+5 to +52
STORAGE TEMPERATURE oF/oC -65 to +185/-54 to +85
Accessories Supplied

Instruction manual, and mating connector for supply voltages.


Calibration

Bouche Labs Charge Amplifier Calibration Service No. 11, or Accelerometer Calibration Service No. 7 with Bouche Labs or customer accelerometer calibrated as a system with the Model 2121F Charge Amplifier.

BOUCHE LABS reserves the right to change these and any other specifications without notice.